Using the right way: pipettes
1-Fill the pipette with 5 mm more than the volume mark. Wipe the exterior surface of the pipette with a paper filter.
2-Hold the pipette at he eye level and fix the tip on a slightly tilted beaker wall.
3-Let the exces volume to pour until the liquid meniscus is at the same level with the volume mark.
4-When the liquid get to the tip WAIT accordingly to the waiting time in the product.
5-At the end of the waiting time support and move the tip on the beaker's wall and transfer the liquid.
6-DO NOT BLOW the remaining liquid at the tip of the pipette.