Using the right way: Burettes
Using the right way: Burettes
1-Fill the burette a litle bit more then the "0" level
2-Check the valve - open it and let the liquid to pour but not under the zero level.
3-Check for air bubble in the burette.
4-Fill the burette 10 mm more above the zero level.
5-Set the zero level by checking the meniscus at the eye level.
6-Wipe the liquid drops and start the titration. When the color change appear wait for 30 seconds.
7-Get in the beaker the last drop as this is also part of the titration volume as well.